by David Zielske, MD, CIRCC, COC, CCVTC, CCC, CCS, RCC

Question:
For the following, would it be appropriate to report code 93641 (DFT), or would there need to be additional documentation?

“1: Single-chamber cardioverter defibrillator implantation. The device was attached to the lead and implanted.

Device testing was performed. Two-way communication was established between the device and its programmer; telemetered electrograms and pacing and sensing thresholds were measured.”

Answer:
No. There was no defibrillation documented in this report. These are normal programming items for an ICD, similar to a pacemaker.